Literature DB >> 5132095

Aerosol inoculator for exposure of human volunteers.

P J Gerone, R B Couch, V Knight.   

Abstract

The performance of an aerosol inoculator for human volunteers is described in tests that used the PR8 strain of type A influenza virus and sodium fluorescein as a physical tracer. Virus recovery from the aerosols was approximately 1% and was unaffected by such variables as prolonged aerosolization, total airflow, relative humidity, or method of sampling. The recovery of sodium fluorescein from the aerosol was approximately 12% and was influenced by total airflow rates and relative humidity. With this apparatus, it should be possible to deliver reasonably predictable and measurable doses of respiratory viruses to human subjects. The design makes it possible to dismantle the inoculator into its component parts to facilitate portability.

Entities:  

Mesh:

Substances:

Year:  1971        PMID: 5132095      PMCID: PMC376441          DOI: 10.1128/am.22.5.899-903.1971

Source DB:  PubMed          Journal:  Appl Microbiol        ISSN: 0003-6919


  2 in total

1.  A MOBILE LABORATORY UNIT FOR EXPOSURE OF ANIMALS AND HUMAN VOLUNTEERS TO BACTERIAL AND VIRAL AEROSOLS.

Authors:  W R GRIFFITH
Journal:  Am Rev Respir Dis       Date:  1964-02

2.  An apparatus for the study of airborne infection.

Authors:  D W HENDERSON
Journal:  J Hyg (Lond)       Date:  1952-03
  2 in total
  1 in total

Review 1.  Methods for sampling of airborne viruses.

Authors:  Daniel Verreault; Sylvain Moineau; Caroline Duchaine
Journal:  Microbiol Mol Biol Rev       Date:  2008-09       Impact factor: 11.056

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.