| Literature DB >> 494351 |
U Quast, J F Goncalves, K H Schick, B Strauch, L Glaeser, M Busch.
Abstract
In-vivo-dosimetry is necessary to prevent overdosage during radiation treatment techniques of great risk. New special dosemeter probes with up to five Si-pn-junction-diodes in a silicone tube were developed and tested for endodosimetry. They show high efficiency and good linearity of response, without dose-rate dependence. The steep increase of the sensitivity for low energies causes a phantom depth dependence of response. A low temperature dependence but a large anisotropy of response is found for the tested diodes. These multidetector probes proved to be sufficient for direct reading endodosimetric control of dose and dose-distribution during radiotherapy.Entities:
Mesh:
Substances:
Year: 1979 PMID: 494351
Source DB: PubMed Journal: Strahlentherapie ISSN: 0039-2073