| Literature DB >> 472263 |
Abstract
Radiation exposure to operating room personnel was measured by thermoluminescent dosimeters and an ionization chamber during a simulated iodine-125 seed implantation of the prostate gland. The exposure rates at points around the implanted phantom were measured and compared with the calculated values. The data suggested that even with a schedule of 25 prostate implants per year, exposure levels would not exceed 5% of the recommended maximum limits for occupationally exposed personnel.Entities:
Mesh:
Substances:
Year: 1979 PMID: 472263 DOI: 10.1148/132.3.748
Source DB: PubMed Journal: Radiology ISSN: 0033-8419 Impact factor: 11.105