Literature DB >> 440786

Further observations on noise levels in infant incubators.

F H Bess, B F Peek, J J Chapman.   

Abstract

The purpose of this study was to conduct an acoustic analysis of incubator noise under two conditions: when the incubator was associated with different types of life-support equipment; and when impulse noise was created by striking the side of the incubator or by opening and closing the doors of the storage unit. It was found that the life-support equipment increased the overall noise levels of incubators by as much as 15 to 20 dB. Much of this increased energy was in the high frequency region. Impulse signals created by striking the side of the incubator ranged from 130 to 140 dB. A representative impulse for opening the incubator was 92.8 dB, whereas closing the door produced a peak amplitude of 114 dB.

Entities:  

Mesh:

Year:  1979        PMID: 440786

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  4 in total

1.  Noise levels within the ear and post-nasal space in neonates in intensive care.

Authors:  S S Surenthiran; K Wilbraham; J May; T Chant; A J B Emmerson; V E Newton
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2003-07       Impact factor: 5.747

Review 2.  Evaluation of the child with deafness or delayed speech.

Authors:  J Coplan
Journal:  Indian J Pediatr       Date:  1988 May-Jun       Impact factor: 1.967

3.  Environmental neonatology.

Authors:  D Wolke
Journal:  Arch Dis Child       Date:  1987-10       Impact factor: 3.791

4.  The "Sound of Silence" in a Neonatal Intensive Care Unit-Listening to Speech and Music Inside an Incubator.

Authors:  Matthias Bertsch; Christoph Reuter; Isabella Czedik-Eysenberg; Angelika Berger; Monika Olischar; Lisa Bartha-Doering; Vito Giordano
Journal:  Front Psychol       Date:  2020-05-26
  4 in total