Literature DB >> 4079875

Characteristics of electron beams from a new 25-MeV linear accelerator.

P O'Brien, H B Michaels, J E Aldrich, J W Andrew.   

Abstract

Clinically useful electron fields are produced on the Atomic Energy of Canada, Limited Therac 25 linear accelerator by computer-controlled scanning of the electron beam. Measurements were made to determine the properties of these electron fields. Central axis percentage depth dose and bremsstrahlung background were compared for these fields and for the fields from selected machines that use scattering foils. Dose calibrations were made in both water and polystyrene using the American Association of Physicists in Medicine Task Group 21 protocol. Measurements were made to determine the relative output factors, virtual source position, and the attenuation of the electron fields by lead.

Entities:  

Mesh:

Year:  1985        PMID: 4079875     DOI: 10.1118/1.595788

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Comparison of measured electron energy spectra for six matched, radiotherapy accelerators.

Authors:  David J McLaughlin; Kenneth R Hogstrom; Daniel W Neck; John P Gibbons
Journal:  J Appl Clin Med Phys       Date:  2018-03-30       Impact factor: 2.102