A study of the characteristics of radiation contaminants within a clinically useful photon beam.

G Krithivas, S N Rao.   

Abstract

Contaminant radiation within a therapeutic beam has been studied for accelerator-produced 24-MV x rays by a direct measurement utilizing independent jaws. A carefully positioned diode was exposed to secondary radiation for various collimator settings that project field sizes between [10 X (0 + 5)] and [26 X (0 + 13)] cm at a source-to-axis distance of 1 m. [The notation [L X (W1 + W2)] means the projected field area due to lower (L) and upper (W1, W2) jaws.] Measurements were taken by placing polystyrene sheets with density thicknesses ranging from 0.21 to 4.53 g cm-2 in front of the detector. The data strongly demonstrate that with increasing field size, the ratio of the dose due to electrons to that due to photons increases rapidly. The characteristic feature of the electron depth dose curves may be linked to the observed shift in the depth of dose maximum (dmax) with field size. Data taken with a magnetic field of 0.15 T permit analysis of photons and electrons with regard to their intensity, energy, and behavior in a phantom. From an analysis of 12.5 X 12.5 cm data, various radiation components have been studied and are correlated with the observed contaminants.

Entities:  

Mesh:

Year:  1985        PMID: 4079869     DOI: 10.1118/1.595661

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  1 in total

1.  Surface and buildup dose characteristics for 6, 10, and 18 MV photons from an Elekta Precise linear accelerator.

Authors:  Eric E Klein; Jacqueline Esthappan; Zuofeng Li
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.