| Literature DB >> 4079459 |
M J Borrelli, W G Carlini, W C Dewey, B R Ransom.
Abstract
A simple procedure for manufacturing Cl-, K+, and pH liquid membrane ion-sensitive microelectrodes is presented in detail. Electrodes suitable for recording from the specimen of interest are back-filled with a small amount of silane solution and heated for 5 min on a hot plate at a temperature between 400 and 500 degrees C, after which they are injected with the ion-sensitive resin. The procedure is adaptable to many different glass stocks, e.g., single-barreled, double-barreled, or theta glass, and can be used to produce electrodes having a wide range of tip sizes for recording either extracellular or intracellular ion activities. Another advantage of the method is speed; up to 10 electrodes can be prepared simultaneously, permitting over 40 functional electrodes to be made per hour.Entities:
Mesh:
Year: 1985 PMID: 4079459 DOI: 10.1016/0165-0270(85)90051-2
Source DB: PubMed Journal: J Neurosci Methods ISSN: 0165-0270 Impact factor: 2.390