Literature DB >> 4076433

Ultrasound follicle diameter measurement: an assessement of interobserver and intraobserver variation.

M K Eissa, K Hudson, M F Docker, R S Sawers, J R Newton.   

Abstract

The intraobserver and interobserver variation of ultrasound measurement of Graafian follicles was assessed. Fourteen follicles from 10 to 30 mm in diameter were each measured three times in three orthogonal planes by three observers. The intraobserver standard deviation (SD) was 0.6 mm and interobserver SD, 1.2 mm, irrespective of follicle diameter. The pooled SD for both variables was 1.2 mm, giving 95% confidence limits of +/- 2.4 mm for any measurement.

Mesh:

Year:  1985        PMID: 4076433     DOI: 10.1016/s0015-0282(16)49032-1

Source DB:  PubMed          Journal:  Fertil Steril        ISSN: 0015-0282            Impact factor:   7.329


  1 in total

1.  Comparison of the results of vaginal and abdominal follicle scans.

Authors:  J Deutinger; A Reinthaller; P Riss; G Bernaschek; P Csaicsich; F Fischl; E Müller-Tyl
Journal:  Arch Gynecol Obstet       Date:  1987       Impact factor: 2.344