Comparison of ionisation measurements in water and polystyrene for electron beam dosimetry.

I A Bruinvis, S Heukelom, B J Mijnheer.   

Abstract

For the determination of absorbed dose to water in electron beams, dosimetry protocols advocate ionisation measurements in plastic phantoms instead of water for practical reasons. The chamber readings in polystyrene at the depth of maximum ionisation must be corrected for the difference in physical properties between the two materials. This correction factor was determined for a Farmer 0.6 cm3 graphite-walled chamber in electron beams with mean energies at the phantom surface between 6 and 19 MeV. Experiments with white polystyrene yielded corrections for the measured ionisation ranging from 0.3 to 2.4%. For clear polystyrene, 0.6-1% higher corrections were found. For beams with the same mean energy at the phantom surface, but with different beam-flattening and collimation systems, variations in this correction up to 1.2% were observed. In contrast to recent reports on electrical charge storage in polystyrene due to electron irradiation, our experiments do not show differences in the chamber readings any larger than 0.5%.

Entities:  

Mesh:

Substances:

Year:  1985        PMID: 4070361     DOI: 10.1088/0031-9155/30/10/003

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Evaluation of an innovative plastic cube phantom designed to improve the efficiency of accelerator QA.

Authors:  J C Huang; L E Reinstein
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

2.  Differences in electron beam dosimetry using two commercial ionization chambers and the TG-21 protocol: another reason to switch to TG-51.

Authors:  David S Followill; William F Hanson; Geoffrey S Ibbott; Leon R Eglezopoulos; Chen-Shou Chui
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.