| Literature DB >> 4067553 |
Abstract
This is a report of a lethal venous air embolism complicating percutaneous ultrasonic lithotripsy. Sudden death occurred during the course of the procedure. Evidence of air embolism was detected at autopsy. Air embolism occurred because a suction pump was connected backwards, forcing air into the urinary tract rather than aspirating irrigating fluid as intended. This mishap has not previously been reported as a complication of this procedure.Entities:
Mesh:
Year: 1985 PMID: 4067553
Source DB: PubMed Journal: J Forensic Sci ISSN: 0022-1198 Impact factor: 1.832