Literature DB >> 4062654

Effects of magnetic resonance imaging fields on stapedectomy prostheses.

E L Applebaum, G E Valvassori.   

Abstract

Seven different types of widely used metallic stapedectomy prostheses were individually placed on a millimeter scale in a plastic Petri dish. The Petri dish was moved within the core and around the magnet of a 1.5-tesla magnetic resonance imaging unit (General Electric Sigma). No movement of any of the prostheses was seen. We conclude that there is no apparent danger of these prostheses becoming displaced in stapedectomy patients subjected to the electromagnetic fields of magnetic resonance imaging units.

Entities:  

Mesh:

Year:  1985        PMID: 4062654     DOI: 10.1001/archotol.1985.00800140064012

Source DB:  PubMed          Journal:  Arch Otolaryngol        ISSN: 0003-9977


  2 in total

1.  Postoperative complications in otospongiosis: usefulness of MR imaging.

Authors:  A S Rangheard; K Marsot-Dupuch; A S Mark; B Meyer; J M Tubiana
Journal:  AJNR Am J Neuroradiol       Date:  2001 Jun-Jul       Impact factor: 3.825

2.  Behavior of metal implants used in ENT surgery in 7 Tesla magnetic resonance imaging.

Authors:  Ariane Thelen; Hans-Christian Bauknecht; Patrick Asbach; Thomas Schrom
Journal:  Eur Arch Otorhinolaryngol       Date:  2006-07-12       Impact factor: 2.503

  2 in total