| Literature DB >> 4055947 |
H J Nelis, M F Lefevere, E Baert, W D'Hoore, A P De Leenheer.
Abstract
Chromatographic procedures have been developed for determination of the stabilizers N-acetyl-DL-tryptophan and octanoic acid in human albumin solutions. N-Acetyl-DL-tryptophan and the internal standard, N-formyl-DL-tryptophan, were separated by liquid chromatography on a reversed-phase column with UV detection at 280 nm. Deproteinization and extraction were carried out with methanol. The extraction recovery at the level of 4.9 mM was 92.5 +/- 2.5% (S.D.) (n = 10), and the average coefficient of variation (C.V.) for replicate analyses of albumin solutions (mean = 2.57, 10.44 and 17.10 mM) was 1.10% (n = 27). Octanoic acid was determined gas chromatographically as its methyl ester, with nonanoic acid as the internal standard. The sample pretreatment included acidification, extraction with hexane and derivatization with methanol-sulphuric acid. The relative recovery from albumin solutions was 89.7 +/- 5.8% (S.D.) (n = 6), and replicate determinations of the compound yielded a C.V. of 5.5% (mean = 14.82 mM, n = 9).Entities:
Mesh:
Substances:
Year: 1985 PMID: 4055947 DOI: 10.1016/s0021-9673(01)87367-2
Source DB: PubMed Journal: J Chromatogr