| Literature DB >> 4036881 |
N P Kubasik, T K Mayer, A G Bhaskar, H E Sine, J P D'Souza.
Abstract
The authors evaluated the Kodak Ektachem Slides for total, "direct," conjugated, unconjugated, and albumin-bound ("delta") bilirubin. For various concentrations of control material, the precision (CV) within- and between-day ranged from 1.7 to 2.4% (2.8-6.6%) for total bilirubin, 0.8-4.2% (1.6-11.5%) for unconjugated bilirubin, and 1.6-9.5% (2.6-20.7%) for conjugated bilirubin. The Ektachem total and "direct" bilirubin assays demonstrated excellent correlation with the Jendrassik and Grof procedure; a 30% difference was observed, however, between absolute numbers with the two direct bilirubin methods. We found the measurement of true conjugated bilirubin by the Kodak Method to be superior to the measurement of "direct" bilirubin in following the response to treatment of various hepatic disorders manifested by extrahepatic and intrahepatic biliary obstruction.Entities:
Mesh:
Substances:
Year: 1985 PMID: 4036881 DOI: 10.1093/ajcp/84.4.518
Source DB: PubMed Journal: Am J Clin Pathol ISSN: 0002-9173 Impact factor: 2.493