CT measurement of the diameter of spinal and other bony canals: effects of section angle and thickness.

B A Eubanks, C E Cann, M Brant-Zawadzki.   

Abstract

The effects of section angle and thickness on the measurement of spinal canal diameter by computed tomography (CT) were tested. A phantom of three spinal canals was constructed of three Lucite rods of different diameters, each covered with 1.8-mm-thick Teflon. A second phantom was constructed of two Lucite rods of the same diameter covered with 1.8- and 8.8-mm-thick Teflon, respectively. CT sections 10, 5, and 1.5 mm thick were obtained at various angles relative to the transverse plane of each canal. Anteroposterior diameters were measured at bone window settings using a standard software package and a method that uses a CT-number profile. Results show that sections obtained at an angle to the transverse plane of the canal do not always overestimate its diameter, as previously suggested, but can make it appear artifactually stenotic. The effect is from volume averaging and is most marked when thick, severely angled sections are used.

Entities:  

Mesh:

Year:  1985        PMID: 4034968     DOI: 10.1148/radiology.157.1.4034968

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

1.  Lumbar spinal canal morphometry from computed tomography scans: reproducibility, results and clinical implications.

Authors:  C G Greenough; R D Fraser
Journal:  Eur Spine J       Date:  1992-06       Impact factor: 3.134

2.  Segmental spinal canal volume in patients with degenerative spondylolisthesis.

Authors:  Jun Miao; Shaobai Wang; Won Man Park; Qun Xia; Xiutong Fang; Martin P Torriani; Kirkham B Wood; Guoan Li
Journal:  Spine J       Date:  2013-03-27       Impact factor: 4.166

3.  Impact of pedicle-lengthening osteotomy on spinal canal volume and neural foramen size in three types of lumbar spinal stenosis.

Authors:  P Li; L Qian; W D Wu; C F Wu; J Ouyang
Journal:  Bone Joint Res       Date:  2016-06       Impact factor: 5.853

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.