Dependence of magnetic resonance image (MRI) intensity values on relaxation times, pulse intervals and other signal attenuation factors.

W S Yamanashi, D W Anderson, P D Lester, D Herrick, J W Frazer.   

Abstract

Magnetic resonance image (MRI) pixel intensities were investigated using a phantom containing several uniform size chambers filled with solutions of known relaxation times, as well as head scans of patients and volunteers. Intensities were measured with a variety of pulse intervals typically used for imaging with spin echo, (SE) and inversion recovery (IR) sequences at 0.15 Tesla using the back projection (R-THETA) method, and at 0.27 Tesla using the 2-dimensional Fourier transform (2DFT) technique. The results were compared with the calculated dependence of MRI signal intensity on relaxation times and pulse interval parameters using the well known functions containing exponential forms. The experimental and the calculated pixel intensity time dependence did not always agree. We infer that factors other than the conventional functions for T1 and T2 signal decay are important. These factors may include the attenuation of the radiofrequency (RF) signals through inhomogenious lossy dielectric materials (e.g., tissues and organs), the location (coordinate) of the portion of the sample to be imaged relative to the RF coils, and the timing and amplitude of gradient pulses relative to the RF input and the detected signals. The flow velocity and diffusions are also important determinants of the signal from blood vessels and body fluids. We point out the necessity for further investigation toward more comprehensive understanding of MRI intensities.

Entities:  

Mesh:

Year:  1985        PMID: 4034683

Source DB:  PubMed          Journal:  Physiol Chem Phys Med NMR        ISSN: 0748-6642


  1 in total

1.  The signal intensity of preoperative magnetic resonance imaging has predictive value for determining the arthroscopic reparability of the anterior talofibular ligament.

Authors:  Jungtae Ahn; Jung Gyu Choi; Bi O Jeong
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2020-08-08       Impact factor: 4.342

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.