Literature DB >> 4033345

Evaluation of the carbon dioxide laser for performance of gastrointestinal myotomies.

J LoCicero, J W Frederiksen, R S Hartz, M W Kaufman, L L Michaelis.   

Abstract

Using a conventional carbon dioxide laser set at 8-W continuous-wave energy, a technique was developed in the animal laboratory to incise only the muscularis of the pylorus leaving the mucosa and submucosa intact. Similar esophageal myotomies were performed with less success, due to the rich submucosal venous plexus causing obscuring hemorrhage and to the reapproximation of the longitudinal esophageal muscle over a 3-week period. The technique for esophageal myotomy requires further refinement, but pyloromyotomy appears applicable for controlled human trials.

Entities:  

Mesh:

Year:  1985        PMID: 4033345     DOI: 10.1002/lsm.1900050409

Source DB:  PubMed          Journal:  Lasers Surg Med        ISSN: 0196-8092            Impact factor:   4.025


  1 in total

Review 1.  New uses of the laser in thoracic and cardiovascular surgery.

Authors:  L L Michaelis; J LoCicero; R S Hartz; W J McCarthy
Journal:  Jpn J Surg       Date:  1990-11