Literature DB >> 4032105

Parametric analyses of pin-bone stresses in external fracture fixation devices.

R Huiskes, E Y Chao, T E Crippen.   

Abstract

Problems occurring in the use of external fixators for bone fractures include pin-bone interface necrosis, infection, and loosening. These may be initiated and enhanced by pin-bone interface stress levels. Based on stress data from finite element method (FEM) models, an analytical "closed-form" model of the local pin-bone configuration in long-bone fracture fixation is developed. This model is relatively simple and useful for routine applications in combination with clinical studies and animal experiments. Although approximate, the most significant stress predictions in the pin-bone structure compare well with more sophisticated FEM analysis results. The analytical model is used for extensive parametric analyses, investigating the effects on the pin-bone interface stress distribution of frame configuration parameters, pin diameter and modulus, bone dimensions, and elastic characteristics. The results indicate that these stresses may reach very high levels under unfavorable circumstances but can be drastically reduced by increasing the bending rigidity of the pin, reducing the side-bar separation and applying full-pin configuration in favor of half-pins.

Entities:  

Mesh:

Substances:

Year:  1985        PMID: 4032105     DOI: 10.1002/jor.1100030311

Source DB:  PubMed          Journal:  J Orthop Res        ISSN: 0736-0266            Impact factor:   3.494


  7 in total

1.  Drug loading and release of Tobramycin from hydroxyapatite coated fixation pins.

Authors:  Mirjam Lilja; Jan Henrik Sörensen; Ulrika Brohede; Maria Astrand; Philip Procter; Jörg Arnoldi; Hartwig Steckel; Maria Strømme
Journal:  J Mater Sci Mater Med       Date:  2013-06-19       Impact factor: 3.896

2.  Low-magnitude high-frequency mechanical signals accelerate and augment endochondral bone repair: preliminary evidence of efficacy.

Authors:  Allen E Goodship; Timothy J Lawes; Clinton T Rubin
Journal:  J Orthop Res       Date:  2009-07       Impact factor: 3.494

3.  Functional and radiological outcome in management of compound tibia diaphyseal fracture with AO monolateral fixator versus Limb reconstruction system.

Authors:  Hitesh J Mangukiya; Neetin P Mahajan; Eknath D Pawar; Aakash Mane; Jitsen Manna
Journal:  J Orthop       Date:  2018-01-31

4.  Stability with unilateral external fixation in the tibia.

Authors:  N Giotakis; B Narayan
Journal:  Strategies Trauma Limb Reconstr       Date:  2007-04

5.  Time-dependent behaviour of bone accentuates loosening in the fixation of fractures using bone-screw systems.

Authors:  S Xie; K Manda; P Pankaj
Journal:  Bone Joint Res       Date:  2018-11-03       Impact factor: 5.853

6.  RELATIONSHIP BETWEEN RIGIDITY OF EXTERNAL FIXATOR AND NUMBER OF PINS: COMPUTER ANALYSIS USING FINITE ELEMENTS.

Authors:  Marcelo Back Sternick; Darlan Dallacosta; Daniela Águida Bento; Marcelo Lemos do Reis
Journal:  Rev Bras Ortop       Date:  2015-11-04

7.  Development of a Novel in Silico Model to Investigate the Influence of Radial Clearance on the Acetabular Cup Contact Pressure in Hip Implants.

Authors:  Saverio Affatato; Massimiliano Merola; Alessandro Ruggiero
Journal:  Materials (Basel)       Date:  2018-07-25       Impact factor: 3.623

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.