E Sveinsdottir, B Larsen, P Rommer, N A Lassen.
Abstract
A computer-based scintillation camera has been designed for both dynamic and static radionuclide studies. The detecting head has 254 independent sodium iodide crystals, each with a photomultiplier and amplifier. In dynamic measurements simultaneous events can be recorded, and 1 million total counts per second can be accommodated with less than 0.5% loss in any one channel. This corresponds to a calculated deadtime of 5 nsec. The multidetector camera is being used for 133Xe dynamic studies of regional cerebral blood flow in man and for 99mTc and 197 Hg static imaging of the brain.Entities:
Mesh:
Substances:
Year: 1977 PMID: 401876
Source DB: PubMed Journal: J Nucl Med ISSN: 0161-5505 Impact factor: 10.057