| Literature DB >> 401460 |
Abstract
Various whole-body, multiple-field, electron-beam therapy techniques are analyzed. Accelerator characteristics, various dosimetric techniques, and experimental setup are described. A flattening filter was constructed which produced an uncollimated electron-beam profile (4 MeV) with a uniformity of +/- 8% across a 200-cm diameter area at 7 m from the source. Irradiation techniques include 2-, 4-, 6-, and 8-field therapy and rotation. Conclusive evidence is presented to support adoption of either the 8-field technique or rotation as the methods of choice for superficial whole-body treatments.Mesh:
Year: 1977 PMID: 401460 DOI: 10.1148/122.1.219
Source DB: PubMed Journal: Radiology ISSN: 0033-8419 Impact factor: 11.105