Literature DB >> 4005680

Use of a urologic set for improved fluid administration rates.

S Nadeau, M Tousignant.   

Abstract

This study was designed to compare flow rates of three sets of tubings: a large bore (8 mm) Y-urologic set (Y-URO); a Y-URO with an added in-line blood filter (Y-URO + BF); and a currently used (3.5 mm diameter) blood transfusion set (Y-TRANSFUSION). The time required to empty a one litre bag of Ringer's Lactate from a 1.0 meter vertical drop was measured while using four different IV catheters (9.5, 10, 14 and 16 gauge), and the flow rates calculated. When a 9.5-gauge catheter was used, the flow rates obtained by the Y-URO and the Y-URO + BF were respectively 269 and 162 per cent higher than those achieved by the Y-TRANSFUSION. This wide margin was decreased to 41 and 39 per cent with a 16-gauge catheter. A 9.5-gauge catheter, when compared to a 16-gauge, allowed a 346 per cent improvement in flow rates with the Y-URO, a 225 per cent with the Y-URO + BF and only a 71 per cent increase with the Y-TRANSFUSION. The urologic set allows better fluid flow because of the larger diameter of its tubing. This adaptation of a large bore urologic irrigation set can be very useful when rapid fluid administration is imperative.

Entities:  

Mesh:

Year:  1985        PMID: 4005680     DOI: 10.1007/bf03015144

Source DB:  PubMed          Journal:  Can Anaesth Soc J        ISSN: 0008-2856


  3 in total

1.  Clinical significance of blood viscosity.

Authors:  J A Dormandy
Journal:  Ann R Coll Surg Engl       Date:  1970-10       Impact factor: 1.891

2.  Pressurized infusion system for fluid resuscitation.

Authors:  J H Philip; B K Philip
Journal:  Anesth Analg       Date:  1984-08       Impact factor: 5.108

3.  Flow rate variance of commonly used IV infusion techniques.

Authors:  D J Dula; H A Muller; J W Donovan
Journal:  J Trauma       Date:  1981-06
  3 in total
  1 in total

1.  Evaluation of 3-mm diameter intravenous tubing for the rapid infusion of fluids.

Authors:  G D Cross
Journal:  Arch Emerg Med       Date:  1987-09
  1 in total