Servo-controlled air pump for calibration of respiratory measurement systems.

G Saumon, A Loiseau, E Delavault.   

Abstract

We describe a servo-controlled piston pump driven by a stepping motor. The analogue controller is a single non-linear second-order feedback loop with adjustable speed and acceleration limits. This system, designed to simulate slowly-moving, active and non-linear systems, can be used as a low (0-6.5 litre s-1) flow volume generator in calibration procedures.

Mesh:

Year:  1985        PMID: 3999723     DOI: 10.1016/0141-5425(85)90042-1

Source DB:  PubMed          Journal:  J Biomed Eng        ISSN: 0141-5425


  1 in total

1.  Linear servo-controlled pressure generator for forced oscillation measurements.

Authors:  P L de Melo; M M Werneck; A Giannella-Neto
Journal:  Med Biol Eng Comput       Date:  1998-01       Impact factor: 2.602