Underestimation of prosthetic mitral valve areas: role of transseptal catheterization in avoiding unnecessary repeat mitral valve surgery.

M H Schoenfeld, I F Palacios, A M Hutter, S S Jacoby, P C Block.   

Abstract

In patients with symptoms of heart failure after mitral valve replacement, identification of a stenosed prosthesis may be difficult. Twelve such patients were evaluated, presenting at a mean of 8.4 years after mitral valve replacement (four mechanical, eight porcine). Transvalvular pressure gradients were obtained using both indirect (pulmonary capillary wedge) and direct (transseptal catheterization) measurements of left atrial pressure. In all 12 patients, the diastolic gradient across the prosthetic valve was overestimated when pulmonary wedge rather than transseptal measurements were used. Calculated mitral valve prosthetic area was underestimated by the pulmonary wedge determinations. These findings may be caused by either the phase delay of the pulmonary wedge V wave relative to the transseptal V wave, resulting in a higher diastolic mean left atrial pressure, or the faulty wedge determinations in the setting of pulmonary hypertension, or both. In patients being considered for repeat mitral valve replacement because of prosthetic valve stenosis, transseptal catheterization allows for more accurate determination of prosthetic valve area and more accurately defines the need for repeat mitral valve surgery.

Entities:  

Mesh:

Year:  1985        PMID: 3998319     DOI: 10.1016/s0735-1097(85)80353-3

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  2 in total

1.  Correlation of pulmonary capillary wedge pressure with left atrial pressure in patients with mitral stenosis undergoing balloon valvotomy.

Authors:  Desabandhu Vinayakumar; Uppalakal Bijilesh; C G Sajeev; Gopalan Nair Rajesh; Cicy Bastion; P Kadermuneer; V Haridasan; Dolly Mathew; Biju George; M N Krishnan
Journal:  Indian Heart J       Date:  2016-01-18

Review 2.  Quantitative applications of Doppler cardiography in congenital heart disease.

Authors:  S D Colan
Journal:  Cardiovasc Intervent Radiol       Date:  1987       Impact factor: 2.740

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.