A comparative computational study of blood flow through prosthetic heart valves using the finite element method.

S R Idelsohn, L E Costa, R Ponso.   

Abstract

The steady flow of blood through three common types of prosthetic heart valves was simulated numerically using the finite element method. The velocity, pressure and stress fields were obtained for the disk-type, tilting-disk and ball-type prosthetic heart valves in aortic position, for increasing Reynolds numbers up to 900, 1500 and 2000 respectively. Computer graphics of calculated velocities are presented, showing in detail the accelerated flow, recirculation and stagnation areas developed in the prosthesis. Maximum wall shear stresses were found at 0.5, 1.4, and 1.2 diameters from the sewing ring downstream for the disk, tilting-disk and ball valves being the values of 55, 18 and 33 dyn cm-2 respectively. In the vicinity of the occluder, maximum shear stresses of 38, 30 and 47 dyn cm-2 respectively were computed. The flow characteristics and performance for each valve are compared, the results are presented in terms of energy loss and maximum shear stress. The velocity and stress fields are compared with in vitro evaluations found in the literature.

Mesh:

Year:  1985        PMID: 3988789     DOI: 10.1016/0021-9290(85)90003-x

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  2 in total

1.  Numerical study of turbulent blood flow through a caged-ball prosthetic heart valve using a boundary-fitted co-ordinate system.

Authors:  K Thalassoudis; J Mazumdar; B J Noye; I H Craig
Journal:  Med Biol Eng Comput       Date:  1987-03       Impact factor: 2.602

2.  Role of computational fluid mechanics in the analysis of prosthetic heart valve flow.

Authors:  G D Tansley; R J Edwards; C R Gentle
Journal:  Med Biol Eng Comput       Date:  1988-03       Impact factor: 2.602

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.