Effects of scattered X-rays on CT images.

H Kanamori, N Nakamori, K Inoue, E Takenaka.   

Abstract

Scatter to primary radiation ratios at detector positions were calculated for uniform cylindrical water phantoms in both third and fourth generation x-ray CT scanners using the Monte Carlo simulation method. A rapid method for determining the scattering angle is given. In the process of image reconstruction, the effect of scatter was isolated by applying our new calculation technique. The decrease in CT number due to scatter was found to be less than 10. In the third generation CT scanner the rate of scatter is larger at the centre than at the edge because of the effect of the collimator at the detectors. As a result the cupping artefact occurs in the CT image. The collimator cannot be used in the fourth generation CT scanner and, therefore, the rate of scatter is almost uniform, leading to the appearance of rounded-off edges.

Entities:  

Mesh:

Year:  1985        PMID: 3983234     DOI: 10.1088/0031-9155/30/3/004

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  7 in total

1.  Assessment of errors caused by X-ray scatter and use of contrast medium when using CT-based attenuation correction in PET.

Authors:  Mohammad Reza Ay; Habib Zaidi
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-04-19       Impact factor: 9.236

Review 2.  Current status and new horizons in Monte Carlo simulation of X-ray CT scanners.

Authors:  Habib Zaidi; Mohammad Reza Ay
Journal:  Med Biol Eng Comput       Date:  2007-07-05       Impact factor: 2.602

3.  An Accurate Scatter Measurement and Correction Technique for Cone Beam Breast CT Imaging Using Scanning Sampled Measurement (SSM) Technique.

Authors:  Xinming Liu; Chris C Shaw; Tianpeng Wang; Lingyun Chen; Mustafa C Altunbas; S Cheenu Kappadath
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2006-02-28

Review 4.  Artefacts in CBCT: a review.

Authors:  R Schulze; U Heil; D Gross; D D Bruellmann; E Dranischnikow; U Schwanecke; E Schoemer
Journal:  Dentomaxillofac Radiol       Date:  2011-07       Impact factor: 2.419

5.  CT Hounsfield numbers of soft tissues on unenhanced abdominal CT scans: variability between two different manufacturers' MDCT scanners.

Authors:  Ramit Lamba; John P McGahan; Michael T Corwin; Chin-Shang Li; Tien Tran; J Anthony Seibert; John M Boone
Journal:  AJR Am J Roentgenol       Date:  2014-11       Impact factor: 3.959

6.  Accuracy of volumetric bone mineral density measurement in high-resolution peripheral quantitative computed tomography.

Authors:  Kiranjit Sekhon; Galateia J Kazakia; Andrew J Burghardt; Bryan Hermannsson; Sharmila Majumdar
Journal:  Bone       Date:  2009-06-06       Impact factor: 4.398

7.  Evaluation of algebraic iterative image reconstruction methods for tetrahedron beam computed tomography systems.

Authors:  Joshua Kim; Huaiqun Guan; David Gersten; Tiezhi Zhang
Journal:  Int J Biomed Imaging       Date:  2013-05-27