| Literature DB >> 3976838 |
T Grosvenor, D M Perrigin, J Perrigin.
Abstract
On the basis of routine clinical findings taken on a group of 100 patients serving as subjects in the Houston Myopia Control study, a three-way comparison was made between retinoscopy, subjective, and Dioptron Nova refractive findings. Retinoscopy and subjective findings compared very favorably, mean differences being -0.01 D for spherical equivalent power and -0.05 D for cylinder power. Dioptron and retinoscopy findings differed by -0.30 D for spherical equivalent power and -0.13 D for cylinder power, whereas Dioptron and subjective findings differed by -0.33 D for spherical equivalent power and -0.18 D for cylinder power.Entities:
Mesh:
Year: 1985 PMID: 3976838 DOI: 10.1097/00006324-198501000-00009
Source DB: PubMed Journal: Am J Optom Physiol Opt ISSN: 0093-7002