The ability of an HPLC gel exclusion column (TSK G6000PW) to separate lipid vesicles, viruses, and biological vesicles according to size was tested and compared with separations on Sephacryl S1000. The columns were calibrated using vesicular Stokes radii determined by quasielastic light scattering. The vesicles separated according to size on both types of column and remained intact during elution. Viruses of known diameters and clathrin-coated vesicles were also eluted as a function of size. The TSK G6000PW column was able to separate larger particles (greater than 500 nm) than the Sephacryl S1000, and, when used in combination with the TSK G5000PW column, gave more discrete separations of smaller particles (10 to 30 nm diameter). Moreover, the HPLC columns can be run significantly faster (10-20 min vs several hours) and give more precise results than Sephacryl S1000. Therefore, HPLC using a G6000PW column alone, or in combination with a G5000PW column, provides a rapid and accurate means of sizing and selecting specifically sized biological and artificial vesicles.Entities: