Literature DB >> 3947167

Pacemaker backup-mode reversion and injury during cardiac surgery.

G A Lamas, E M Antman, J P Gold, N S Braunwald, J J Collins.   

Abstract

Both electrocautery and internal defibrillation are routinely used in cardiac operations. The cases of 5 patients are presented in whom backup-mode reversion or injury of permanently implanted pulse generators occurred during a cardiac procedure. The theoretical explanations for backup-mode reversion and generator or tissue injury are discussed, and recommendations are made for the management of patients with a pacemaker who are to undergo a cardiac operation.

Entities:  

Mesh:

Year:  1986        PMID: 3947167     DOI: 10.1016/s0003-4975(10)62658-6

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  4 in total

Review 1.  [Perioperative management of patients with implanted pacemakers or cardioverter/defibrillators. Recommendations of the Austrian Society for Anaesthesiology, Resuscitation and Intensive Care Medicine, the Austrian Society for Cardiology and the Austrian Society for Surgery].

Authors:  H Gombotz; M Anelli Monti; N Leitgeb; M Nürnberg; B Strohmer
Journal:  Anaesthesist       Date:  2009-05       Impact factor: 1.041

Review 2.  Evaluation and Monitoring of Patients With Cardiovascular Implantable Electronic Devices Undergoing Noncardiac Surgery.

Authors:  Eduard Sklyar; Jonathan N Bella
Journal:  Health Serv Insights       Date:  2017-03-20

3.  Inappropriate defibrillator shock during gynecologic electrosurgery.

Authors:  Rachel M Kaplan; Jeremiah Wasserlauf; Rachel H Bandi; Elizabeth M Lange; Bradley P Knight; Susan S Kim
Journal:  HeartRhythm Case Rep       Date:  2018-03-26

Review 4.  Anaesthetic consideration in patients with cardiac implantable electronic devices scheduled for surgery.

Authors:  Murali Chakravarthy; Dattatreya Prabhakumar; Antony George
Journal:  Indian J Anaesth       Date:  2017-09
  4 in total