Evaluation of a transcutaneous oxygen and carbon dioxide monitor in a neonatal intensive care department.

P D Wimberley, P S Frederiksen, J Witt-Hansen, S G Melberg, B Friis-Hansen.   

Abstract

Transcutaneous-PO2 (tc-PO2 (tc-PCO2) at 44 degrees C and transcutaneous-PCO2) at 38, 42, 43 and 44 degrees C were measured with a currently available monitoring system (TCM222, Radiometer, Copenhagen) in 64 newborn infants with severe respiratory insufficiency during the first five days of life. Tc-PCO2 at all four temperatures correlated better with arterial blood-PCO2 (aB-PCO2), than tc-PO2 with aB-PO2. However, the sensitivity and specificity of tc-PO2 and tc-PCO2 were similar with regard to maintaining aB-PO2 and aB-PCO2 within specified limits. Tc-PCO2 increased relatively with increasing electrode temperature by a factor which was similar to the anaerobic temperature coefficient of PCO2 in blood. The coefficient of variation of duplicate measurements was 10% for tc-PO2 and 5% for tc-PCO2. Electrode drift after an average of 3 hours patient monitoring was 2% +/- 6% (1 SD) for tc-PO2 and -3% +/- 6% for tc-PCO2. We conclude that tc-PO2 and tc-PCO2 are a valuable supplement to arterial blood gas measurements, but the variable correlation with arterial blood gas values and the electrode drifts which may occur, mean that they cannot fully replace arterial sampling.

Entities:  

Mesh:

Substances:

Year:  1985        PMID: 3923776     DOI: 10.1111/j.1651-2227.1985.tb10983.x

Source DB:  PubMed          Journal:  Acta Paediatr Scand        ISSN: 0001-656X


  5 in total

Review 1.  Transcutaneous carbon dioxide monitoring.

Authors:  J M Rennie
Journal:  Arch Dis Child       Date:  1990-04       Impact factor: 3.791

2.  Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor.

Authors:  C Morgan; S J Newell; D A Ducker; J Hodgkinson; D K White; C J Morley; J M Church
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1999-03       Impact factor: 5.747

3.  Transorally obtained oxygen tension as an indicator of arterial oxygen tension.

Authors:  S E Opper; E E Fibuch; G F Tuohy
Journal:  J Clin Monit       Date:  1988-04

4.  The reproducibility of the response of the human newborn to CO2 measured by rebreathing and steady-state methods.

Authors:  G Cohen; D J Henderson-Smart
Journal:  J Physiol       Date:  1994-04-15       Impact factor: 5.182

5.  Transcutaneous carbon dioxide and oxygen tension in newborn infants: reliability of a combined monitor of oxygen tension and carbon dioxide tension.

Authors:  S Fanconi; H Sigrist
Journal:  J Clin Monit       Date:  1988-04
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.