Literature DB >> 3919249

Radiation protection aspects of a new high-energy linear accelerator.

P O'Brien, H B Michaels, B Gillies, J E Aldrich, J W Andrew.   

Abstract

The Therac-25 is a new 25-MeV linear accelerator manufactured by Atomic Energy of Canada, Ltd. The first two units have recently been installed in Toronto, Ontario and Halifax, Nova Scotia. Calculations and measurements of primary and secondary radiation levels were made. Neutron dose-equivalent rates were measured inside and outside the room. The maximum leakage rate at 1 m from the accelerator target was 0.4% Sv per peak photon Gy. The tenth value layer for neutrons from the Therac-25, at the entrance to a one-legged maze was found to be 5.5 cm of polyethylene. Measurements were done to estimate daily technologist exposure due to induced activity in the treatment room.

Entities:  

Mesh:

Year:  1985        PMID: 3919249     DOI: 10.1118/1.595793

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Radiation burden of assistant medical technicians at a medical accelerator.

Authors:  M Gründel; F Güthoff
Journal:  J Med Phys       Date:  2008-10

2.  Evaluation of equivalent dose from neutrons and activation products from a 15-MV X-ray LINAC.

Authors:  Isra Israngkul-Na-Ayuthaya; Sivalee Suriyapee; Phongpheath Pengvanich
Journal:  J Radiat Res       Date:  2015-08-11       Impact factor: 2.724

3.  Evaluation of protective equipment for the reduction of radiation exposure to physicians performing fluoroscopically guided lumbar transforaminal epidural steroid injections: A randomized controlled trial.

Authors:  Won-Joong Kim; Seung Hee Yoo; Hahck Soo Park
Journal:  Medicine (Baltimore)       Date:  2020-07-31       Impact factor: 1.817

  3 in total