Uniformity correction for SPECT using a mapped cobalt-57 sheet source.

B E Oppenheim, C R Appledorn.   

Abstract

As little as 1% nonuniformity of the scintillation camera can produce disturbing artifacts in SPECT images. To correct for nonuniformity, acquired images must be divided by the flood-field image of a uniform sheet source. The refillable 99mTc sheet source requires meticulous preparation before each use, and is subject to spillage. We propose the use of a permanent 57Co solid sheet source in conjunction with a map of the radioactivity distribution in the source. The flood-field image of the cobalt source is converted by the map into the equivalent of a flood-field image of an almost perfectly uniform source. An accurate and reproducible method is described for generating a map of the cobalt sheet source from images of it made in multiple positions with the scintillation camera. The mapping process is independent of sensitivity variations across the face of the camera.

Entities:  

Mesh:

Substances:

Year:  1985        PMID: 3872354

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  2 in total

Review 1.  Myocardial single-photon emission computed tomographic quality assurance.

Authors:  J M Baron; P Chouraqui
Journal:  J Nucl Cardiol       Date:  1996 Mar-Apr       Impact factor: 5.952

2.  Evaluation of external beam hardening filters on image quality of computed tomography and single photon emission computed tomography/computed tomography.

Authors:  Nivedita Rana; Dinesh Rawat; Madan Parmar; Devinder Kumar Dhawan; Ashok Kumar Bhati; Bhagwant Rai Mittal
Journal:  J Med Phys       Date:  2015 Oct-Dec
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.