Literature DB >> 3838524

Algorithm for the correction of artifacts.

R C Zeimer, N P Blair, J G Cunha-Vaz.   

Abstract

The sources of vitreous fluorophotometry artifacts are described. An algorithm to define this "spread function" is presented. The use of improved instrumentation and compensation for this artifact by data processing of the "correction" algorithm can improve the results of vitreous fluorophotometry. This algorithm was carried out in 40 eyes and was found to have an error of 0.7 mg/ml at 3 mm from the choroid-retina.

Mesh:

Substances:

Year:  1985        PMID: 3838524     DOI: 10.1007/bf02133679

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  4 in total

1.  Vitreous fluorophotometry for clinical research. II. Method of data acquisition and processing.

Authors:  R C Zeimer; N P Blair; J G Cunha-Vaz
Journal:  Arch Ophthalmol       Date:  1983-11

2.  Studies on the technique of vitreous fluorophotometry.

Authors:  R C Zeimer; J G Cunha-Vaz; M E Johnson
Journal:  Invest Ophthalmol Vis Sci       Date:  1982-05       Impact factor: 4.799

3.  Fluorophotometry. I. Technique.

Authors:  R Klein; J T Ernest; R L Engerman
Journal:  Arch Ophthalmol       Date:  1980-12

4.  Vitreous fluorophotometry: identification of sources of variability.

Authors:  T C Prager; D J Wilson; G D Avery; J H Merritt; C A Garcia; G Hopen; R E Anderson
Journal:  Invest Ophthalmol Vis Sci       Date:  1981-12       Impact factor: 4.799

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.