Blood viscosity--a comparative study on three rotational instruments.

E Ernst, C Monshausen, A Matrai.   

Abstract

Three rotational viscometers for blood viscometry were compared at moderate and high shear. At high shear, coefficients of variation were all around 1% increasing 5-fold (Wells-Brookfield) and 3-fold (Carri-Med) when shear was lowered. Using pathological samples the coefficients of variation for the Carri Med mashine was doubled. Only the Contraves LS-30 viscometer showed neither an increasing coefficient of variation with declining shear, nor problems when measuring blood with suspension instability.

Mesh:

Year:  1985        PMID: 3834954     DOI: 10.3233/bir-1985-22602

Source DB:  PubMed          Journal:  Biorheology        ISSN: 0006-355X            Impact factor:   1.875