Umbilical catheters and arterial blood pressure monitoring.

S R Goodwin, S A Graves, J van der Aa.   

Abstract

The natural frequencies, damping coefficients, and accuracies of umbilical artery catheters were determined. The damping coefficients for the 3.5, 5.0, and 8.0 French catheters were 0.40 +/- 0.04 (mean +/- SD), 0.42 +/- 0.05, and 0.19 +/- 0.02, respectively. The natural frequencies were 24.2 +/- 3.2 Hz (mean +/- SD), 18.4 +/- 3.5 Hz, and 26.8 +/- 2.9 Hz, respectively. Measurements obtained with 3.5 and 8.0 French catheters were within 6% of the reference pressure at all pressures and rates tested. With the 5.0 French catheter, however, error greater than 10% from the reference pressure occurred when the rate was 200 pulses per minute or greater and the applied maximum pressure was 100 mm Hg or more.

Mesh:

Year:  1985        PMID: 3831265     DOI: 10.1007/bf02832815

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  3 in total

1.  Physiologic recording by modern instruments with particular reference to pressure recording.

Authors:  D L FRY
Journal:  Physiol Rev       Date:  1960-10       Impact factor: 37.312

2.  Damped and undamped frequency responses of underdamped catheter manometer systems.

Authors:  G G Shapiro; L J Krovetz
Journal:  Am Heart J       Date:  1970-08       Impact factor: 4.749

3.  Direct blood pressure measurement--dynamic response requirements.

Authors:  R M Gardner
Journal:  Anesthesiology       Date:  1981-03       Impact factor: 7.892

  3 in total
  1 in total

1.  Dynamic response of a neonatal catheter-manometer system in situ.

Authors:  H van Langen; P Brienesse; K Kopinga; P Wijn
Journal:  J Clin Monit       Date:  1993-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.