| Literature DB >> 3818858 |
E Rajakylä, K Laasasenaho, P J Sakkers.
Abstract
An high-performance liquid chromatographic method is described for determination of deoxynivalenol, patulin, diacetoxyscirpenol, HT-2-toxin, T-2-toxin, zearalenone and ochratoxin A using a reversed-phase column and a diode-array detector. The extraction and purification steps and optimum chromatographic conditions are described. Detection limits and recoveries from spiked wheat samples were investigated. The combination of the high-performance liquid chromatographic system described together with a modern thermospray quadrupole mass spectrometer is a very specific and sensitive method for analyzing a wide range of mycotoxins in biological samples.Entities:
Mesh:
Substances:
Year: 1987 PMID: 3818858 DOI: 10.1016/s0021-9673(01)94686-2
Source DB: PubMed Journal: J Chromatogr