Abstract
Both isocratic and gradient elution systems for fluorometric detection of K vitamins after post-column reduction with zinc metal to their hydroquinones are described. The reaction detection system for K vitamins (phylloquinone and menaquinones) in liquid chromatography is based on reduction of K vitamins to their corresponding hydroquinones with zinc metal in the presence of zinc ions. It was found that 95% of the injected quinones (K vitamins) could be reduced to their corresponding hydroquinones with zinc metal compared to 60% reduction for electrochemical detectors. Menaquinones could be detected down to 100 pg with relative ease during gradient elution.Entities:
Mesh:
Substances:
Year: 1987 PMID: 3818857 DOI: 10.1016/s0021-9673(01)94685-0
Source DB: PubMed Journal: J Chromatogr