Literature DB >> 3789373

Respiratory monitoring with a new impedance plethysmograph.

S Hoffman, R Jedeikin, D Atlas.   

Abstract

A new impedance plethymograph respiratory monitor, (AR-8800), was tested on anaesthetised dogs. Decreases in tidal volume and total apnoea were brought about by partial and complete airway obstruction and by partial and total muscle paralysis. The changes in respiration were reliably detected by the monitor and its alarms activated within 15-20 seconds. The misinterpretation of gross body movement as normal breathing did not occur. Our findings suggest that the AR-8800 is a safe and dependable monitor which has distinct advantages and may be of particular use in the monitoring of patients at risk of developing respiratory depression or obstruction, whether in the intensive care unit, the general ward or at home.

Entities:  

Mesh:

Year:  1986        PMID: 3789373     DOI: 10.1111/j.1365-2044.1986.tb12966.x

Source DB:  PubMed          Journal:  Anaesthesia        ISSN: 0003-2409            Impact factor:   6.955


  2 in total

Review 1.  Critical review of non-invasive respiratory monitoring in medical care.

Authors:  M Folke; L Cernerud; M Ekström; B Hök
Journal:  Med Biol Eng Comput       Date:  2003-07       Impact factor: 2.602

2.  Comparative provocation test of respiratory monitoring methods.

Authors:  Mia Folke; Fredrik Granstedt; Bertil Hök; Håkan Scheer
Journal:  J Clin Monit Comput       Date:  2002-02       Impact factor: 2.502

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.