| Literature DB >> 3784994 |
R C Speiser, E M Zanrosso, L S Jeromin, R A Carlson.
Abstract
Dose estimates were made for Kodak Min-R screens combined with Kodak Min-R film and Kodak Ortho M film, both with and without a 5:1 Bucky grid; for standard xeroradiographic techniques in negative development mode; and for the new, higher sensitivity xeroradiographic process of the Xerox 175 System. The estimates were derived from exposure versus depth measurements in phantoms made of BR12 breast simulation material using thermoluminescent detectors. A molybdenum target source with molybdenum filtration, at a half-value layer of 0.37-mm Al, was used for the screen-film measurements. All xeroradiographic measurements were made with a tungsten target source with aluminum filtration at half-value layers of 1.5 to 1.56 mm Al. Mean glandular dose estimates for the Min-R screen/Ortho M film combination with Bucky grid and for the new xeroradiographic process were found to be similar. Dose reduction with the new xeroradiographic system was achieved through a more sensitive photoreceptor and more sensitive development, which also improved the unique imaging characteristics of xeroradiography.Entities:
Mesh:
Year: 1986 PMID: 3784994 DOI: 10.1118/1.595871
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071