Catheter separation during cardiac catheterization and coronary angiography. A report of four incidents and review of the literature.

Z S Kyriakides, I P Bellenis, D G Caralis.   

Abstract

Four incidents of catheter separation in three patients during left heart catheterization and coronary angiography are presented. In two cases the catheters had been used before. After the second incident, we decided to use angiography catheters once only. All catheter fragments were removed under local anesthesia, and no further complications arose. In all incidents the point of catheter fracture was near the bond between the body of the catheter and its tip. We believe that catheter separation is related to polymer aging. The catheters in the third and fourth incidents were unused, but there was no expiration date on their packages. We recommend that expiration dates be clearly stated on every catheter package, catheter manufactures issue warnings not to use catheters after expiration date, and unused angiography catheters with no expiration date not be used. We also suggest that consideration be given to using catheters once only.

Entities:  

Mesh:

Year:  1986        PMID: 3767067     DOI: 10.1177/000331978603701010

Source DB:  PubMed          Journal:  Angiology        ISSN: 0003-3197            Impact factor:   3.619


  2 in total

1.  Paradoxical embolisation of a catheter fragment to a coronary artery in an infant with congenital heart disease.

Authors:  H S Klys; A P Salmon; J V De Giovanni
Journal:  Br Heart J       Date:  1991-10

2.  Foreign body associated non-ST elevation myocardial infarction: a case report.

Authors:  Pooyan Dehghani; Mahdi Sajedi Khanian
Journal:  J Cardiovasc Thorac Res       Date:  2017-12-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.