Abstract
A comprehensive study was performed using the full-frame bit-allocation compression technique on 78 radiological images, including digitized radiographs, computed radiographs, and computed tomography images. Each radiograph was digitized to 2,048 X 2,048 X 10, 1,024 X 1,024 X 10, and 512 X 512 X 10 matrices, respectively, with a laser scanner. Five compression ratios were used to compress each image, and reconstructed images from each compressed data set were obtained. Altogether, the authors studied 842 images, including the original and compressed-reconstructed images and the images obtained from the difference between the original and the reconstructed images. The results indicate that acceptable compression ratios for 2,048, 1,024, and 512 matrices are 25:1, 20:1, and 10:1, respectively, based on a mean-square error of 0.02%.Entities: