Literature DB >> 3729619

Fracture of the Carpentier-Edwards ring in tricuspid position: a report of three cases.

M Galiñanes, J Duarte, D F de Caleya, D Garcia-Dorado, F Fdez-Avilés, L M Elbal.   

Abstract

Three cases of fracture of a Carpentier-Edwards ring in tricuspid position are presented. In all three cases the shape of the ring was shown to be abnormal by roentgenogram examination; there was an abrupt bend in the middle portion of each ring. At the time of operation, fracture of the ring skeleton at this midpoint was found in all three instances. Roentgenogram examination seems to be a reliable method of diagnosing this uncommon entity, and tricuspid valve replacement is required for its correction.

Entities:  

Mesh:

Year:  1986        PMID: 3729619     DOI: 10.1016/s0003-4975(10)61840-1

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  3 in total

Review 1.  Flexible band versus rigid ring annuloplasty for tricuspid regurgitation: a systematic review and meta-analysis.

Authors:  Nelson Wang; Steven Phan; David H Tian; Tristan D Yan; Kevin Phan
Journal:  Ann Cardiothorac Surg       Date:  2017-05

2.  Comparison of outcomes of tricuspid annuloplasty with 3D-rigid versus flexible prosthetic ring for functional tricuspid regurgitation secondary to rheumatic mitral valve disease.

Authors:  Haiping Wang; Xiancheng Liu; Xin Wang; Zhenqian Lv; Xiaojun Liu; Ping Xu
Journal:  J Thorac Dis       Date:  2016-11       Impact factor: 2.895

Review 3.  Is a rigid tricuspid annuloplasty ring superior to a flexible band when correcting secondary tricuspid regurgitation?

Authors:  Tie-Yuan Zhu; Jian-Gang Wang; Xu Meng
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-08-20
  3 in total