Literature DB >> 3724695

A precise method for correcting count-rate losses in scintillation cameras.

M T Madsen, R J Nickles.   

Abstract

Quantitative studies performed with scintillation detectors often require corrections for lost data because of the finite resolving time of the detector. Methods that monitor losses by means of a reference source or pulser have unacceptably large statistical fluctuations associated with their correction factors. Analytic methods that model the detector as a paralyzable system require an accurate estimate of the system resolving time. Because the apparent resolving time depends on many variables, including the window setting, source distribution, and the amount of scattering material, significant errors can be introduced by relying on a resolving time obtained from phantom measurements. These problems can be overcome by curve-fitting the data from a reference source to a paralyzable model in which the true total count rate in the selected window is estimated from the observed total rate. The resolving time becomes a free parameter in this method which is optimized to provide the best fit to the observed reference data. The fitted curve has the inherent accuracy of the reference source method with the precision associated with the observed total image count rate. Correction factors can be simply calculated from the ratio of the true reference source rate and the fitted curve. As a result, the statistical uncertainty of the data corrected by this method is not significantly increased.

Mesh:

Year:  1986        PMID: 3724695     DOI: 10.1118/1.595874

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Single cell ganglioside catabolism in primary cerebellar neurons and glia.

Authors:  David C Essaka; Jillian Prendergast; Richard B Keithley; Ole Hindsgaul; Monica M Palcic; Ronald L Schnaar; Norman J Dovichi
Journal:  Neurochem Res       Date:  2012-03-11       Impact factor: 3.996

2.  Six orders of magnitude dynamic range in capillary electrophoresis with ultrasensitive laser-induced fluorescence detection.

Authors:  Colin D Whitmore; David Essaka; Norman J Dovichi
Journal:  Talanta       Date:  2009-08-05       Impact factor: 6.057

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.