A laboratory investigation of two new portable gas analysers.

A M Zbinden, D Westenskow, D A Thomson, B Funk, J Maertens.   

Abstract

Two new portable infrared gas analysers, the Irina (Drägerwerke, Lübeck, Germany) and the Normac (Datex Instrumentarium Corporation, Helsinki, Finland), were tested and the results compared to those from such established methods as gaschromatography and the Beckman LB-2 (Beckman, USA) infrared gas analyser using exactly defined gas mixtures from a vaporizer developed in our laboratory. The analysers were evaluated for their accuracy and precision, noise, zero stability, gain stability and the impact of flow, over-pressure, carrier gas and humidity. All three analysers showed good accuracy and precision. The noise level was acceptable except in the older version of the Normac, but this error has now been corrected by the manufacturer. Zero and gain stability are very good; gas flow and over-pressure do not affect the measurement. The influence of carrier gases and humidity is negligible except for the strong N2O effect on the Beckman. We conclude that this new generation of highly sophisticated and reliable but handy analysers will meet the clinical demands.

Entities:  

Mesh:

Substances:

Year:  1986        PMID: 3722981     DOI: 10.1007/bf02915882

Source DB:  PubMed          Journal:  Int J Clin Monit Comput        ISSN: 0167-9945


  5 in total

1.  Closed-loop control for anesthesia breathing systems.

Authors:  D R Westenskow; C F Wallroth
Journal:  J Clin Monit       Date:  1990-07

2.  Influence of ventilatory and circulatory changes on the pharmacokinetics of halothane and isoflurane.

Authors:  F J Frei; D A Thomson; A M Zbinden
Journal:  Experientia       Date:  1988-02-15

3.  Self-calibrating interference refractometer using a linear photodiode array for anaesthetic gas analysis.

Authors:  R K Jackson; E Palayiwa; B R Sugg; C E Hahn; S Weller; W L Davies
Journal:  Med Biol Eng Comput       Date:  1988-09       Impact factor: 2.602

4.  Accuracy and cross-sensitivity of 10 different anesthetic gas monitors.

Authors:  B Walder; R Lauber; A M Zbinden
Journal:  J Clin Monit       Date:  1993-11

5.  Continuous monitoring of alveolar and inspiratory concentrations of anesthetic and respiratory gases is difficult and potentially unsafe.

Authors:  M K Sykes
Journal:  J Clin Monit       Date:  1987-04
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.