Collection of blood on filter paper for measurement of glycated hemoglobin by affinity chromatography.

R R Little, E M McKenzie, H M Wiedmeyer, J D England, D E Goldstein.   

Abstract

We present data on a filter-paper collection and assay method for measurement of glycated hemoglobin (gHb). Onto filter paper dipped into a solution of glucose oxidase (EC 1.1.3.4) and allowed to dry, approximately 20 microL of capillary blood was spotted. For analysis, we eluted the dried blood spot from the paper by soaking in water for 1 h, then measured the gHb in the eluate by affinity chromatography. Because the gHb significantly increased from day 1 to day 14 of storage, it was necessary to standardize the day of elution from the paper. We found a high correlation between gHb measured from samples stored for 14 days on treated paper and gHb measured by affinity chromatography from frozen whole blood or hemolysates (r = 0.96). This method is convenient, requiring small amounts of blood and little sample handling and assay time, and may be particularly useful in certain situations such as large-scale screening for diabetes.

Entities:  

Mesh:

Substances:

Year:  1986        PMID: 3698280

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  7 in total

1.  Analysis of hemoglobin A1c from dried blood spot samples with the Tina-quantR II immunoturbidimetric method.

Authors:  Trevor G Jones; Kimbrough D Warber; Billy D Roberts
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

2.  Measurement of HbA1c in Gingival Crevicular Blood Using a High-Pressure Liquid Chromatography Procedure.

Authors:  Michael A Pesce; Shiela M Strauss; Mary Rosedale; Jane Netterwald; Hangli Wang
Journal:  Lab Med       Date:  2015

Review 3.  What a drop can do: dried blood spots as a minimally invasive method for integrating biomarkers into population-based research.

Authors:  Thomas W McDade; Sharon Williams; J Josh Snodgrass
Journal:  Demography       Date:  2007-11

4.  Measurement of glycated hemoglobin A1c from dried blood by turbidimetric immunoassay.

Authors:  Ramakrishnan Lakshmy; Ruby Gupta
Journal:  J Diabetes Sci Technol       Date:  2009-09-01

5.  The use of dried blood spot sampling for the measurement of HbA1c: a cross-sectional study.

Authors:  Claudio A Mastronardi; Belinda Whittle; Robert Tunningley; Teresa Neeman; Gilberto Paz-Filho
Journal:  BMC Clin Pathol       Date:  2015-07-08

6.  Comparability of HbA1c and lipids measured with dried blood spot versus venous samples: a systematic review and meta-analysis.

Authors:  Eshan T Affan; Devarsetty Praveen; Clara K Chow; Bruce C Neal
Journal:  BMC Clin Pathol       Date:  2014-05-12

7.  HbA1c determination from HemaSpot™ blood collection devices: comparison of home prepared dried blood spots with standard venous blood analysis.

Authors:  J M Hall; C F Fowler; F Barrett; R W Humphry; M Van Drimmelen; S M MacRury
Journal:  Diabet Med       Date:  2019-10-17       Impact factor: 4.359

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.