Comparison of calibration methods for respiratory inductive plethysmography in infants.

M D Revow1, S J England, H A Stogryn, D L Wilkes.   

Abstract

In infants under the age of 6 mo respiratory inductive plethysmograph (RIP)-calculated tidal volumes (VT) were compared with simultaneously measured volumes using a pneumotachograph (PNT) to 1) assess whether using multiple points (MP) along the inspiratory profile of a breath is superior to using only VT when calculating volume-motion (VM) coefficients, 2) verify the assumption of independent contributions of the abdomen and rib cage to VT, which was accomplished by extending the normal RIP model to include a term representing interaction between these two compartments, and 3) investigate whether VM coefficients are sleep-state dependent. Neither use of multiple points nor inclusion of the interacting term improved the performance of the RIP over that observed using a simple two-compartment model with VT measurements. However, VM coefficients obtained during quiet sleep (QS) were not reliable when used during rapid-eye-movement (REM) sleep, suggesting that coefficients obtained during one sleep state may not be applicable to another state where there is a substantial change in the relative abdominal/rib cage contributions to VT.

Entities:  

Mesh:

Year:  1987        PMID: 3693219     DOI: 10.1152/jappl.1987.63.5.1853

Source DB:  PubMed          Journal:  J Appl Physiol (1985)        ISSN: 0161-7567


  3 in total

1.  The Accuracy of Respiratory Calibration Methods for Estimating Lung Volume During Speech Breathing: A Comparison of Four Methods Across Three Adult Cohorts.

Authors:  Victoria S McKenna; Jessica E Huber
Journal:  J Speech Lang Hear Res       Date:  2019-07-22       Impact factor: 2.297

2.  Reducing the airflow waveform distortions from breathing style and body position with improved calibration of respiratory effort belts.

Authors:  Tiina M Seppänen; Olli-Pekka Alho; Tapio Seppänen
Journal:  Biomed Eng Online       Date:  2013-09-28       Impact factor: 2.819

3.  Respiratory inductance plethysmography calibration for pediatric upper airway obstruction: an animal model.

Authors:  Robinder G Khemani; Rutger Flink; Justin Hotz; Patrick A Ross; Anoopindar Ghuman; Christopher J L Newth
Journal:  Pediatr Res       Date:  2014-10-03       Impact factor: 3.756

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.