We have developed a beat-to-beat calculator which can calculate in real-time the ratio of the diastolic pressure time index (DPTI), and the tension time index (TTI) as an index of the myocardial oxygen supply/demand balance. Physicians set up presumed value for the left ventricular endodiastolic pressure, a search area for the dicrotic notch, a threshold for the onset of the up-slope and the corresponding value of the calibration signal on the digital switches of the calculator. Next, the arterial pressure analog signal is input into the calculator. The calculator searches automatically for both the onset of the up-slope and the dicrotic notch. The arterial pressure curve is displayed beat-to-beat with the recognized onset and the dicrotic notch on the CRT to be confirmed by physicians. When physicians do not agree with the automatic recognition they can fit the automatic recognition to the observation. If the recognition of the onset is inadequate, the threshold can be re-adjusted to trigger the onset. If recognition of the dicrotic notch is inadequate, the physician can adjust the search-area. Therefore, physicians who operate the calculator can rely on the calculated DPTI/TTI. This calculator can continuously monitor the myocardial oxygen supply/demand balance in patients with acute myocardial infarction or just after open-heart surgery.Entities: