| Literature DB >> 3667832 |
K E Choi1, J A Sinkule, D S Han, S C McGrath, K M Daly, R A Larson.
Abstract
A sensitive and specific high-performance liquid chromatographic (HPLC) assay was developed for the quantitation of mitoxantrone in plasma using electrochemical detection. Bisantrene was chosen as the internal standard. A reversed-phase, 10-microns muBondapak C18 analytical column (30 cm X 3.9 mm) with an isocratic mobile phase of 28% acetonitrile in 80 mM sodium formate buffer (pH 3.0) was used. The eluent was monitored by both electrochemical detection at an applied potential of +0.75 V vs. Ag/AgCl and visible absorbance at 660 nm. Only electrochemical detection was able to quantitate the internal standard and provided ten times higher sensitivity than visible absorbance for mitoxantrone with a detection limit as low as 0.1 ng/ml. Calibration curves in the range 0.1-1000 ng/ml showed good linearity (r = 0.998) and precision (coefficient of variation less than 10%). This HPLC method utilized a reproducible and inexpensive liquid-liquid extraction procedure. Using methylene chloride, the extraction efficacy of mitoxantrone from plasma was 85.3% with a coefficient of variation less than 2.1%. This new assay was then applied to measure mitoxantrone concentrations in plasma obtained from two leukemic patients receiving 12 mg/m2 mitoxantrone as a 1-h infusion.Entities:
Mesh:
Substances:
Year: 1987 PMID: 3667832 DOI: 10.1016/0378-4347(87)80157-3
Source DB: PubMed Journal: J Chromatogr