Carbon dioxide retention associated with a humidifier defect.

S McNulty1, L Barringer, J Browder.   

Abstract

A case is presented which describes a patient who developed hypercarbia resulting from a defective humidifier. A Puritan-Bennett Cascade I humidifier was incorporated into a circle system between the inspiratory dome one way valve and the patient. A screw which supports the thermal well to the head of the humidifier was missing, thereby allowing an intermittent leak to develop in the system. The leak was present when the system pressure dropped to ambient level; however, at positive system pressure the leak sealed. This allowed exhaled gases into the inspiratory limb of the circuit unchecked by the inspiratory valve yet when the breathing circuit was occluded at the patient end and submitted to pressures of 20 and 40 cm H2O, no leaks were detected. Intraoperatively, PaCO2 was noted to be as high as 68 mmHg just prior to removing the humidifier from the circuit and corrected to 38 mmHg within 15 minutes of removal of the humidifier.

Entities:  

Mesh:

Year:  1987        PMID: 3664921     DOI: 10.1007/BF03014362

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  6 in total

1.  Rebreathing during controlled respiration with various semiclosed anesthetic systems.

Authors:  D J WATERS; W W MAPLESON
Journal:  Br J Anaesth       Date:  1961-08       Impact factor: 9.166

2.  Carbon dioxide retention using a humidified "nonrebreathing" system.

Authors:  J A Nimocks; J H Modell; P A Perry
Journal:  Anesth Analg       Date:  1975 Mar-Apr       Impact factor: 5.108

Review 3.  Humidification in anaesthesia. A review of the present situation.

Authors:  J E Boys; T H Howells
Journal:  Br J Anaesth       Date:  1972-08       Impact factor: 9.166

4.  Effects of dry anesthetic gases on tracheobronchial ciliated epithelium.

Authors:  J Chalon; D A Loew; J Malebranche
Journal:  Anesthesiology       Date:  1972-09       Impact factor: 7.892

5.  Preventable anesthesia mishaps: a study of human factors.

Authors:  J B Cooper; R S Newbower; C D Long; B McPeek
Journal:  Anesthesiology       Date:  1978-12       Impact factor: 7.892

6.  Humidity and the anesthetized patient.

Authors:  J Chalon; C Patel; M Ali; S Ramanathan; L Capan; C K Tang; H Turndorf
Journal:  Anesthesiology       Date:  1979-03       Impact factor: 7.892

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.