Literature DB >> 3657263

Accuracy of sizers for aortic valve prostheses.

L I Bonchek1, M W Burlingame, B E Vazales.   

Abstract

Accurate sizing of the aortic anulus is a prerequisite of prosthetic valve implantation, because it is usually desirable to insert the largest possible prosthesis. Although it is generally assumed that commercial valve sizers are accurately labeled, our measurements reveal that the sizers for the St. Jude Medical, Björk-Shiley, and Ionescu-Shiley prostheses are larger than their nominal sizes. These findings are relevant to the selection of prosthetic valves.

Mesh:

Year:  1987        PMID: 3657263

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  1 in total

1.  Preoperative aortic annulus size assessment by transthoracic echocardiography compared to the size of surgically implanted aortic prostheses.

Authors:  I J Ingimarsdóttir; L Hellgren Johansson; F A Flachskampf
Journal:  Echo Res Pract       Date:  2019-06-01
  1 in total