Radiation dose to the patient and personnel was determined during extracorporeal shock wave lithotripsy treatment of 60 patients. Surface radiation dose to the patient's back from the fluoroscopy unit on the side with the kidney stone averaged 10 rem (100 mSv.) per case, although the range was wide (1 to 30 rem). The surface dose from the opposing biplane x-ray unit was less, averaging 5.5 rem (55 mSv.) per case but again with a wide range (0.1 to 21 rem). Exit dose at the lower abdomen averaged 13 mrem. (0.13 mSv.) per case and estimated female gonad dose averaged 100 mrem. (1.2 mSv.). Radiation dose to personnel working in the extracorporeal shock wave lithotripsy suite averaged less than 2 mrem. (0.02 mSv.) per case, making it a procedure that is safe in regard to radiation exposure.

Entities:  

Mesh:

Year:  1987        PMID: 3656517     DOI: 10.1016/s0022-5347(17)43350-7

Source DB:  PubMed          Journal:  J Urol        ISSN: 0022-5347            Impact factor:   7.450