Measurement of beam current and evaluation of scatter production in an 18-MeV accelerator.

D L Watts, G S Ibbott.   

Abstract

Measurements of dose rate in a 10-MV x-ray beam may indicate considerable variation with collimator setting. This variation is observed even when the peak scatter factor is fixed by holding the field size at the phantom surface constant. The resulting output factor variation may be due to radiation backscattered from the collimators into the monitor chamber. Measurements of charge deposited in the accelerator target were compared with ionization readings for several field sizes. Evaluation of the data indicates that radiation backscattered into the monitor chamber contributes only a small amount to the output factor.

Mesh:

Year:  1987        PMID: 3627007     DOI: 10.1118/1.596036

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  Prediction of back-scatter radiations to a beam monitor chamber of medical linear accelerators by use of the digitized target-current-pulse analysis method.

Authors:  Yusuke Suzuki; Naoki Hayashi; Hideki Kato; Hiroshi Fukuma; Yasujiro Hirose; Makoto Kawano; Yoshio Nishii; Masaru Nakamura; Takashi Mukouyama
Journal:  Radiol Phys Technol       Date:  2012-10-25

2.  Measurement of back-scattered radiation from micro multileaf collimator into the beam monitor chamber from a dual energy linear accelerator.

Authors:  K R Muralidhar; P Narayana Murthy; N V N M Sresty; Pramod Kumar Dixit; Rajneesh Kumar; A K Raju
Journal:  J Med Phys       Date:  2007-04